Search TechnoSource site  

        Electraedge

        Output range upto 300V, 400 mA

Cat. no. 28-T0550
with timer


        - Independent digital displays for mA and volts
        - Dual mode (CV and CC) with auto crossover
        - Output V & I can be set to any desired value within range
        - Pause resume function
        - Quick and easy to set timer
        - Four sets of recessed output terminals
        - Compatible with most electrophoresis instruments
        - Can be used in a cold room or incubator
        - Noise, vibration and heat free (fanless design)
        - Can be run for days continuously without any problems
        - Auto-recovery on power failure
        - Light weight aluminium body with handles provided for portability
Download Electraedge brochure pdf




Manual for programming
the Timer pdf

      Safety Features
        - Outputs protected against short-circuiting
        - Over-voltage trip circuit to limit electronic
          component damage
        - High insulation resistance and isolation

      Technical Specifications
FeatureElectraedge (with timer)
28-T0550
  Voltage Output   0-500V
  Current output   0-500mA DC
  Power (wattage)   250 W
  Voltage adjustment   1 V increment
via ten turn knob
  Current adjustment   1 mA increment
via one turn knob
  Mode of regulation   CVCC with LED indication
  Timer   Yes
  Timer Display   Dual, 3 digit LED display
  Time ranges
  Select between nine time ranges
0 - 9.99/ 99.9/ 999 sec
0 - 9.59 min:sec, 99.9/ 999 min
0 - 9.59 hr:min, 99.9/ 999 hr
  Programmable   Yes
  Autorecovery on power failure
  Set point stays indefinitely
  Yes
  Timer recovery on power failure   Yes, memory retention for 10 years
  Pause resume function key   Yes
  Operating temperature range   0°C to +50°C
  Voltage input   230V rms, 50 Hz, single phase
  Voltage Variation (from nominal AC input voltage)   +10%, -15%
  Frequency   47Hz to 53Hz
  Transient suppression provided   Yes
  Component details
  Mains switch
  Fuse
  Pause/ resume switch
  Timer ON/ OFF switch
  Timer reset switch
  Output socket
  Mains inlet
 
  Illuminated rocker switch
  On front panel
  Rocker switch
  Rocker switch
  Push button (momentary)
  4mm recessed female
  IEC C14
  Output is fully isolated from input   Yes
  Line regulation
  in CV mode
  in CC mode
  Better than
+/-0.25%
+/-0.5%
  Load regulation in:
- no load to full load variation (in CV mode)
- change in zero to max output (in CC mode)
 
+/-0.25%
+/-0.5%
  Ripple & noise
  in Cv mode
  in CC mode
 
100mV rms max. 20Hz-20MHz
5mA rms max.
  Automatic overload & short circuit
  protection with self recovery
  Yes
  3.5 digit voltmeter and ammeter
  (independent display)
  Yes
  Isolation
~ between AC input & DC output terminals
~ between DC output terminals and chassis
 
2.1KV AC for 1 min
1.5KV AC for 1 min
  Insulation
~ between output & chassis
 
> 10Mohms @ 250DC
  Humidity tolerance   upto 95% (non condensing)
  Cooling by normal convection, venting at the bottom and sides   Yes
  Dimensions of the body
[measurement excludes projecting knobs, handles and switches]
  160 mm (w) x 250 mm (h) x 255 mm (d)



    For further information about other power supplies please visit

    Features page     and     Selection Guide




Electraedge: Power supply from TechnoSource
For Electrophoresis and Electro-blotting.
This power pack is compatible with most standard Equipment.